Click here for new LABsynergy website which is now live!
Below you will find some useful product details
Independent measuring device for control and evaluation of CO2, temperature and O2
Case
Plastic (ABS)
CO2 measurement
Dual beam infrared sensor Range: 0 - 10 % CO2 Display definition: 0.1 % CO2 Accuracy:
· 0 - 6 % CO2 ± 0.2 % CO2
· 6 % - 10 % CO2 ± 0.3 % CO2
Sample volume per measurement (2 min.): approx. 0.8 L
Temperature measurement
Platinum temperature sensor PT1000 Range: 0 - 100 °C Display definition: 0.1 °C Accuracy:
· 20 °C - 50 °C ± 0.2 °C
· 50 °C - 100 °C ± 0.3 °C
O2 measurement
Galvanic sensor Range: 0 - 100 % O2 Display definition: 0.1 % O2 O2 sensor not included
Data logger
Recording of measurement logging in arbitrary intervals Output of measurements via:
· backlit LCD
· PC software Data Visual’09 (via USB data cable)
Interval for data logging: 15 - 120 min., adjustable in steps of 5 min. Storage of max. 1008 measurements with date and time
Power supply
Rechargeable Li-Ion battery Power supply 100 – 240 V AC, 5 V DC 1.5 A
Operation time
Use of rechargeable battery approx. 4 h Time to recharge battery approx. 4 h
REF 10556: Sensor for O2 measurement