REF 13946
Incubator with an inner volume of 200 liters
Advantages of C200:
- Approved medical device
- Active sterile humidity:
External water reservoir
Distilled water is vaporized at 120 °C
Minimized risk of contamination
Humidity sensor (rH)
Very fast recovery
Dual-beam infrared sensor
Very precise measurement
Very fast recovery
Long-term stable
Independent six-sided direct heating
Homogeneous temperature distribution in the incubation chamber
Very fast recovery
O2 control
Subdivided inner glass door with 6 small doors
Lockable front door
Access port
PC software LaboDat+
Features:
Interior
|
Electro-polished stainless steel with coved corners No tools required for quick disassembly and assembly of interior components
|
Temperature
|
Panel heated chamber and door Intelligent temperature control system avoids condensation in the inner chamber
|
Active sterile Humidity supply
|
The external water supply container
|
CO2 control
|
Dual-beam infrared sensor Measurement independent from temperature and humidity
|
Fast recovery
|
For all adjustable parameters through optimized microprocessor control
|
Easy handling
|
Menu guided Large, backlit LCD Foil keyboard
|
Diagnostic system
|
Optical and acoustic alarm if setpoint deviation or sensor defect occurs Door monitoring Remote alarm
|
Options
|
O2 control: connection for N2 possible Subdivided inner glass door with 6 small doors Lockable front door Access port Ø 30 mm
|
Interfaces
|
Remote alarm (potential free) RS485 interface for connection of optional available PC software LaboDat+
|
Technical Data:
Exterior
|
Dimensions 680 x 690 x 930 mm (w x d x h)
|
Interior
|
Chamber volume 200 l Electro-polished stainless steel Shelves height adjustable
|
Temperature
|
Panel heated interior chamber and door Range 27-42 °C starting 5 °C above ambient temperature Stability / uniformity ±0.1 °C / 0.3 °C Over-temperature alarm 32-47 °C
|
CO2 control
|
Dual-beam infrared sensor Range 0–10 % CO2 Stability ±0.1 % CO2
|
O2 control < 21 % (optional)
|
Measurement with galvanic sensor Connection of N2 Adjustable range 1–21 % O2
|
Humidity
|
Active sterile humidity generated by vaporizing module operating at 120 °C Measuring range 0–98 % rH / adjustable range 60–95 % rH
|
Supply
|
230 V AC, 50/60 Hz or 115 V AC, 50/60 Hz, 200 W Inlet pressure 0.8 bar Ambient temperature 18-30 °C
|
Classification
|
Safety class I
Class IIa for all usage according to EC-Directive 93/42/EEC Conform to EN 61010
|
Options:
- Subdivided inner glass door - for even faster recovery
6 small inner doors
Even faster recovery of all parameters
Even lower gas consumption
Easy cleaning of inner doors
Scope of supply:
-6 split shelves 500 x 222 mm (d x w) each
-3 shelves 502 x 481 mm (d x w) each
- Stacking kit - for safe positioning
Safe, easy and space-saving positioning of the devices in the laboratory
-REF 10649 Underframe, 15 cm height, white coated steel
-REF 10681 High frame stand, 115 cm height, white coated steel
For mobile positioning of the devices in the laboratory, on lockable wheels, preventing roller base movement
REF 14411 Roller base, 15 cm height, on lockable wheels
Ø 30 mm
Located in the upper left corner inside the incubation chamber
Sealing plug in scope of supply
- Lockable front door - higher security for your cultures
Our incubator C200 is available with O2 control as option (O2 < 21 %)
|