REF 14093
Incubator with an inner volume of 60 litres
Advantages of C60:
- Approved medical device
- Active sterile humidity:
External water reservoir
Distilled water is vaporized at 120 °C
Minimized risk of contamination
Humidity sensor (rH)
Very fast recovery
Dual-beam infrared sensor
Very precise measurement
Very fast recovery
Long-term stable
Independent six-sided direct heating
Homogeneous temperature distribution in the incubation chamber
Very fast recovery
O2-control
Subdivided inner glass door with 4 small doors
Lockable front door
Access port
PC software LaboDat+
Features:
Interior
|
Electro-polished stainless steel with coved corners No tools required for quick assembly and disassembly of the interior
|
Temperature
|
Panel heated chamber and door Intelligent temperature control system avoids condensation in the inner chamber
|
Active sterile Humidity supply
|
The external water supply container
|
CO2 control
|
Dual-beam infrared sensor Measurement independent from temperature and humidity
|
Fast recovery
|
For all adjustable parameters through optimized microprocessor control
|
Easy handling
|
Menu guided Large, backlit LCD Foil keyboard
|
Diagnostic system
|
Visual and acoustic alarm if setpoint deviation or sensor defect occurs Door monitoring Remote alarm
|
Options
|
O2 control: connection of N2 Subdivided inner glass door with 4 small doors Lockable front door Access port Ø 30 mm
|
Interfaces
|
Alarm contact (potential free) RS485 interface for connection of optionally available PC software LaboDat+
|
Technical Data:
Exterior
|
Dimensions 510 x 540 x 720 mm (w x d x h)
|
Interior
|
Chamber volume 60 l Electro-polished stainless steel Shelves height adjustable
|
Temperature
|
Panel heated interior chamber and door Range 27-42 °C starting 5 °C above ambient temperature Stability / uniformity ±0.1 °C / ±0.3 °C Over-temperature alarm 32-47 °C
|
CO2 control
|
Dual-beam infrared sensor Range 0–10 % CO2 Stability ±0.1 % CO2
|
O2 control < 21 % (optional)
|
Measurement with galvanic sensor Connection of N2 Adjustable range 1–21 % O2
|
Humidity
|
Active sterile humidity generated by vaporizing module at 120 °C Measuring range 0–98 % rH / range 60–95 % rH
|
Supply
|
230 V AC, 50/60 Hz or 115 V AC, 50/60 Hz, 160 W Inlet pressure for gas 0.8 bar Ambient temperature 18-30 °C
|
Classification
|
Safety class I Class IIa for all usage according to EC-Directive 93/42/EE
Conform to ENN 61010
|
Options:
- Subdivided inner glass door - for even faster recover
4 small inner doors
Even faster recovery of all parameters
Even lower gas consumption
Easy cleaning of inner doors
- Stacking kit - for safe positioning
Safe, easy and space-saving positioning of the devices in the laboratory
REF 10789: High frame stand, 105 cm height, white-coated steel
- Lockable front door - higher security for your cultures
- Access port
Ø 30 mm
Located in the upper right corner of the incubation chamber
Sealing plug in the scope of supply