Labotect CO2 Incubator Labo C201 – IVFSynergy

Labotect CO2 Incubator Labo C201


For more information or a quote on this item please email IVFSynergy

Below you will find some useful product details


REF 15245

Incubator with 200 litre chamber capacity

Advantages of Labo C201:

  • 7“ colour touch screen
  • UV decontamination (option)
  • Particle filter (P3) within the airflow
  • Graphical display of recorded parameters of the past 2 h or 24 h (T, CO2, rH, O2)
  • Cable bushing (Access port)
  • Access control via numerical code (option)
  • RS485 or ethernet interface
  • Active sterile humidity:
External water reservoir
Distilled water is vaporized at 120 °C
Minimized risk of contamination
Humidity sensor (rH)
Very fast recovery
  • CO2sensor:
Dual-beam infrared sensor
Very precise measurement
Very fast recovery
Long-term stable
  • Direct heating:
Independent six-sided direct heating
Homogeneous temperature distribution in the incubation chamber
Very fast recovery
  • Optional:
UV decontamination
O2 control
Subdivided inner glass door with 6 small doors
Access control via numerical code
PC software LaboDat+

Features:

Interior

Electro-polished stainless steel with coved corners
Shelves and ductwork easily removable for cleaning purpose
Access port Ø 30 mm

Temperature

Intelligent temperature control system
avoids condensation in the inner chamber

Humidity supply

Active sterile humidity via an external water supply container

CO2 control

Dual-beam infrared sensor
Measurement independent from temperature and humidity

Fast recovery

For all adjustable parameters through optimized microprocessor control

Easy handling

Menu guided 7“ colour touch display
Graphical display of recorded
parameters of the past 2 h or 24 h

Diagnostic system

Visual and acoustic alarm if the setpoint deviation or sensor defect occurs
Door monitoring
Remote alarm

Options

Subdivided inner glass door with 6 small doors

Access control via numerical code 

UV decontamination routine

PC software LaboDat+

O2 control: connection for N2

Interfaces

Alarm contact (potential free)
RS485 or Ethernet interface

 

Technical Data:

Exterior

Dimensions 720 x 700 x 870 mm (w x d x h)

Interior

Chamber volume 200 l
Electro-polished stainless steel
Shelves height adjustable

Temperature

Panel heated interior chamber and door
Range 27-42 °C starting 5 °C above ambient temperature
Stability / uniformity ±0.1 °C / ±0.3 °C
Over-temperature alarm 32-47 °C

CO2 control

Dual-beam infrared sensor
Range 0–10 % CO2
Stability ±0.1 % CO2

O2 control < 21 % (optional)

Measurement with galvanic sensor
Connection of N2
Range 1-21 % O2

Humidity

Active sterile humidity generated by vaporization module at 120 °C
Measuring range 0–98 % rH / adjustment range 60 % – 95 % rH

Access port

Ø 30 mm

Supply

230 V AC, 50/60 Hz or 115 V AC, 50/60 Hz, 250 VA
Inlet pressure for gas 0.8 bar
Ambient temperature 18-30 °C

Classification

Safety class I
Conform to EN 61010

 

Options:

  • Subdivided inner glass door - for even faster recovery

 6 small inner doors
Even faster recovery of all parameters
Even lower gas consumption
Easy cleaning of inner doors
Scope of supply:

-6 split shelves 474 x 222 mm (d x w) each

-3 shelves 485 x 454 mm (d x w) each

  •  UV decontamination
Individual start by user 
Duration: 180 minutes
Short-wave UV radiation (UV-C radiation, <280 nm)
UV mode is terminated immediately when the door is opened 

 

  • Access control - higher security for your cultures
Access control via touch screen
Mechanical unlocking in case of power failure (emergency release)
  • Stacking kit - for safe positioning
Safe, easy and space-saving positioning of the devices in the laboratory


-REF 10649 Underframe, 15 cm height, white coated steel

-REF 10681 High frame stand, 115 cm height, white coated steel

  •  Roller base

 For mobile positioning of the devices in the laboratory, on lockable wheels, preventing roller base movement

-REF 14411 Roller base, 15 cm height, on lockable wheels

 

Our Labo C201 incubators are available with O2-control (O2 < 21 %)

 Each Labotect CO2 incubator passes an individual cell-growth test prior to delivery!

 

C201_Brochure.pdf