REF 16666 - (without heating block)
Transport incubator for a secure and mobile preservation of specimen, culture media etc.
Features:
- Secure transport of cells without electric mains and CO2 connection
- Factory-aligned parameters: 37 °C and 5 % CO2
- Different temperature or CO2 concentration possible (to be advised before the order)
- Operation at 12 V car voltage possible
- Use of different heating blocks
- Easy handling and cleaning
- Compact Design
Parameters* during transport without connection to electric mains and CO2 bottle:
Time
|
Temp.
|
CO2
|
0 min
|
37 °C
|
5 %
|
10 min
|
36.6 °C
|
4.9 %
|
30 min
|
35.5 °C
|
4.8 %
|
60 min
|
34.0 °C
|
4.7 %
|
*The measurements relate to an ambient temperature of 23°C
Technical Data:
Power supply
|
115 V or 240 V AC, 50/60 Hz 12 V DC with car voltage via adapter cable
|
Dimensions
|
Outside: 385 x 245 x 305 mm (l x w x h) Inside: approx. 312 x 167 x 140 mm (l x w x h)
|
Weight
|
Approx. 4.7 kg
|
Temperature
|
Available between 32 °C and 39 °C (to be advised before the order)
|
Accuracy of regulation
|
± 0.3 °C (measured with heating block REF 10264)
|
Heating up time
|
Approx. 2 h
|
CO2 concentration
|
5 % (different concentrations possible) The heating block inside necessary Connection to CO2 supply with pre-pressure of 0.8 bar necessary CO2 accuracy ± 0.3 % CO2 (when supplied with CO2 with pre-pressure of 0.8 bar ± 0.05 bar)
|
Basic Supply:
- Basic unit
- Mains cable
- Car connection cable